GAMON-S
GAMON-S is a CAEN SyS gamma spectroscopy system for online radiation monitoring indoors and outdoors. Unlike a simple dose rate meter, it identifies isotopes automatically and calculates isotope-related dose rates, helping operators tell natural background variation apart from artificial radionuclides. Several scintillation crystals are available: LaBr3(Ce) at 1.5 x 1.5 in or 2 x 2 in, NaI(Tl) at 2 x 2 in or 3 x 3 in, and CeBr3 at 1.5 x 1.5 in or 2 x 2 in. The scintillator covers 30 keV to 3 MeV, while a Geiger-Mueller detector extends dose rate measurement from 10 nSv/h to 10 Sv/h. The MCA has 2048 channels as standard, or 8192 on request. Power consumption is under 4 W, it operates from -40 to 60 °C, and it is sealed to IP68. Length is 91.5 cm, diameter 14.0 to 17.0 cm, and weight up to 8 kg. It suits environmental monitoring networks, uranium and mineral sands mining, and nuclear medicine facilities. Speak to ProDigitek about detector choice and a quote.

Key specifications
Scintillation crystals | LaBr3(Ce) 1.5 in x 1.5 in or 2 in x 2 in; NaI(Tl) 2 in x 2 in or 3 in x 3 in; CeBr3 1.5 in x 1.5 in or 2 in x 2 in |
Energy range (scintillator) | 30 keV to 3 MeV |
Dose rate range (Geiger-Mueller) | 10 nSv/h to 10 Sv/h |
MCA depth | 2048 channels (8192 on request) |
Power consumption | < 4 W |
Operating temperature | -40 to 60 °C |
Protection level | IP68 |
Size and weight | Length 91.5 cm, diameter 14.0 to 17.0 cm, up to 8 kg |